Cytometry and Antibody Technology

Cytek Aurora

Spectral benchtop analyzers

Home » Benchtop Analyzers » Cytek Aurora

Overview

Model: Aurora
Manufacturer: Cytek
Configuration

The CAT Facility has two identical Cytek Auroras in the main facility, which have been named Aurora Red and Aurora Blue to distinguish them. The PME Aurora is also managed by the CAT Facility, but located in the PME space. All Auroras have a number of exciting features:

  • Spectral flow cytometer
  • 5 lasers, 64 fluorophore detectors, 3 FSC/SSC detectors
  • Detect a wide variety of fluorescent antibodies/dyes/proteins
  • High sensitivity
  • Cope with autofluorescence with the autofluorescence extraction feature
  • Load samples from standard FACS tubes or a variety of plates
  • Volumetric counter allows for cell count
Cytek Aurora in the Cytometry and Antibody Technology Facility

Training Requirements

In order to use any cytometer in the CAT Facility without staff assistance, all users must complete the basic training outlined in the New User Checklist on the New Users page. The training for the Aurora specifically is:

  1. Flow Basics course (schedule online)
  2. Aurora training course
  3. Aurora hands on instrument training
  4. (hands on training for any other benchtop analyzers can be done at any time before you use another cytometer)

The Aurora training course covers all of the basic information needed to plan a successful Aurora experiment – creating proper controls, panel design tools, spectral unmixing, etc. Please contact Laura Johnston to schedule Aurora training.

Sample Preparation Tips

Accepted Tubes

The Auroras accept standard 5mL round bottom tubes (FACS tubes) as well as a variety of 96-well plates, including deep-well. It is possible to run 5mL FACS tubes with small volume tube inserts (AKA bullet tubes), however an extra setup step must be performed on the instrument to prevent damaging the sample introduction port (SIP).

Sample Volume

  • Running samples on the Aurora is comparable to the Fortessas
  • A minimum of 100 µL is reccomended, 150-300 µL is typical but use what is appropriate for the sample
  • It’s best to bring concentrated samples, once you are at the cytometer you can easily dilute your tubes if they are running to quickly
  • In general, run samples at less than 35,000 events/sec

Flow Rate

The volumetric counter allows for live monitoring of the sample flow rate. The estimated rates are:

  • low: 15 µL/min
  • medium: 30 µL/min
  • high: 60 µL/min

Actual samples my vary. For example, many cell types run on low typically have rates around 9-11 µL/min.

Sample Loader

The Aurora-Blue and -Red are equipped with an ASL Plate loader, while the PME Aurora has an AMS Plate Loader. See below for the features on both systems.

Dead Volume

Dead Volume is the volume of the sample remaining inside the sample tube or plate well that cannot be aspirated into sample injection probe after each sample acquisition.

Configuration
Cytek Aurora 5 laser configuration

Troubleshooting

Section under construction, updates coming soon!
How do I move the plate loader out of the way?
Running QC | What to do when QC fails

The SpectroFlo QC beads are available in the black fridge by the entrance of the facility.
– Vortex the bottle of SpectroFlo QC beads 10sec, and add one drop in a 5mL tube. Add 300ul of water.
– Load the tube on the Aurora
– In the software, select QC&Setup, select the Daily QC tab. Make sure the correct bead lot number is selected. Start the acquisition.
– The instrument will make measurements and optimize settings for a few minutes, then a window will appear letting you know if the QC passed or failed.

What to do if QC Fails
If the Aurora won’t even complete the QC run, it may be because the event rate is incorrect (it should be within 100 to 1000 events/sec). It’s likely too low, so vortex the tube of QC beads and add one more drop to the tube.

If QC went through but you get a QC Failed message, review the QC report and identify the problem.

  • The Gain Change has been flagged for one or several channels: this typically occurs if 1 – you used the wrong bead lot number; or 2 – the beads might be too old and have started degrading, in which case you should make a new tube.
  • The %rCV of one or multiple detectors are flagged: a low rCV value is required for proper flow cytometry data. SpectroFlo will only flag the third detector of each laser array when rCVs are too high, even though multiple detectors might be having a similar issue. The common cause for a high rCV will be accumulation of crud on the flow cell that impacts the scatter of the signals.
    • From the left side menu, select the Cytometer icon, and the Clean Flow Cell protocol
    • Ignore the prompts on the screen – run a tube of 30% contrad twice
    • Run the Clean Flow Cell protocol again, but run a tube of water twice this time.
    • Run the QC protocol again, and noticed if the rCV values are going down. Repeat as needed.
    • If the rCV values remain unchanged, the issue could be with the alignment of the lasers themselves. Ask the staff for assistance, do not attempt to align the laser by yourself.

Should you continue with the acquisition if QC fails?
If you can’t get QC to pass on the Aurora, notify the CAT staff asap. We can discuss the next steps. If no one is immediately available and you need to decide if you can proceed with the acquisition on the instrument, consider what the issue is and the impact it may have on your data.

  • rCV values informs how well the instrument will be able to resolve the signals from the different fluorophores. Cytek recommends a rCV value below 6 for most detectors. If you are skirting just above that threshold, and are looking at very bright markers, this may not be a huge issue. Similarly, if the high rCV values come from the detector array of a laser that is not providing particularly useful information for your assay, you could move on with the acquisition. Let me emphasize that this is not the way we want to operate, we’ll want to resolve the problem asap.
I have a low or decreasing event rate
There are no events displayed
How do I clean the instrument at the end of my experiment | at the end of the day

At the end of your experiment:

  • Click on the Aurora icon on the left side of the screen and select Clean Flow Cell
  • Follow the instructions on the screen

 

At the end of the day:

  • Click on the Aurora icon on the left side of the screen and select Fluidic Shutdown
  • Follow the instructions on the screen

 

Data won't transfer to the server
  • The data should upload automatically every 30 minutes, so be sure that you gave enough time for the transfer to occur.
  • Data should b exported to D:/CytekbioExport/FcsFiles/Experiments/FlowUser. Make sure that you did not export your files in another location – either somewhere else on the disk, or a sub-folder of FlowUser.
  • In some case communication gets lost, contact the staff and we will alert our IT manager.
How do I export the Experiment file

From My Experiments in the SpectroFlo software, select your experiment and click Export. Navigate to D:/Experiment Server Transfer. The experiment file will be pushed to the server within 30min.

Contact the staff

Email cathelp@bsd.uchicago.edu to reach the CAT Facility staff.